E.cam or Symbia systems that use foresight detectors - Product Usage: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

E. CAM and Symbia system with foresight detectors performing gated or dynamic acquisition may lose some detector time information. Our risk analysis indicates that the probability of occurrence is remote. SPECT: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. CT: The CT component is intended to produce cross sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial place taken at different angles or spiral planes taken at different angles. SPECT+CT: Perform CT scans and nuclear imaging studies within the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
E.cam or Symbia systems that use foresight detectors - Product Usage: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. — Recall Details · AllClear