FDAJuly 24, 2020device

Siemens Sensis Vibe Hemo system in combination with the MicroPodTM EtCO2 module-diagnostic and administrative tool supporting hemodynamic catheterization and/or electrophysiology studies, for cardiac as well as interventional Radiology.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment conditions, the HemoBox may stop functioning and all vital signs are lost

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical