FDAJuly 24, 2018device

GE 9100c NXT anesthesia systems

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.

What to do

FDA enforcement status: Terminated

Brands named

ge medical systems china

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
GE 9100c NXT anesthesia systems — Recall Details · AllClear