FDAAugust 7, 2024device

Infusomat Space Infusion System/ Large Volume Pump, Product Code 8713050U

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Additional devices have been affected.

What to do

FDA enforcement status: Ongoing

Brands named

b braun medical

UPCs

04046963716752

Recall history