FDAApril 12, 2017device

LASER CDS

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Multiple surgical packs were assembled with a non-sterile CHG applicators labeled as "ChloraPrep" (kit component numbers: 29183, 52115, 82279, and 82281). The non-sterile components 29183, 52115, 82279, and 82281 should have been placed inside the sterile portion of the pack, where it would subsequently be exposed to the ethylene oxide sterilization process. However, the components were not placed in the sterile portion of the pack during assembly and were never exposed to the sterilization process. Components 29183, 52115, 82279, and 82281 were inadvertently released non-sterile. The remainder of the pack components were properly sterilized and only components 29183, 52115, 82279, and 82281 were affected. Components 29183, 52115, 82279, and 82281 are manufactured by Becton, Dickinson (BD) and are labeled "ChloraPrep". ChloraPrep" applicators contain 2% w/v chlorhexidine gluconate (CHG) in 70% v/v isopropyl alcohol (IPA). The applicators are used to facilitate skin preparation for vari

What to do

FDA enforcement status: Terminated

Brands named

medline

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.