FDAAugust 15, 2018device

da Vinci S/Si EndoWrist 8mm Monopolar Curved Scissors Model No. 420179-16 Product Usage: EndoWrist Instruments, including scissors, scalpels, forceps, needle drivers and electrocautery are intended for endoscopic manipulation of tissue, including: grasping, cutting, blunt and sharp dissection, ap...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Monopolar Curved Scissor (MCS) instruments may be susceptible to developing cracks near the proximal edge of the extension tube. These cracks may fracture the plastic surrounding the instrument shaft, and may cause fragments of the extension tube to fall, including potentially into the patient.

What to do

FDA enforcement status: Ongoing

Brands named

intuitive surgicalintuitive

UPCs

00886874111505