FDASeptember 8, 2020device
AC/DC POWER SUPPLY for 7400, Part Number 151004300, XP Power Model Number AHM180PS19-XE0338, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is intended for use with the following portable ultrasound systems: MyLabAlpha (101740000) MyLabGamma ...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.
What to do
FDA enforcement status: Terminated
Brands named
esaote s p aesaote
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAEsaote Endocavity ultrasonic probe, Model E 3-12, REF 1200000382025-03-14
- FDAAC/DC POWER SUPPLY for 7400, Part Number 151004301, XP Power Model Number AHM180PS19-XE1057, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is intended for use with the following portable ultrasound systems: MyLabAlpha (101740000) MyLabGamma ...2020-09-08
- FDAMyLab Gamma, Model 74102018-06-11
- FDAMyLab Alpha, Model 74002018-06-11
- FDAMyLab Seven, Model 64002018-06-11
- FDAMyLab Ultrasound System; Models 6100, 6150, 6200, 6250; PA023E Probe 7.5/102018-01-03
- FDAMyLab Ultrasound System; Models 6100, 6150, 6200, 6250; Probes: BL433, C 18, CA541, EC1123, EC123, IH 618, IOT332, IOT342, L 415, LA332E, LA435, LA523, LA533, SI2C41, SL2325, TRT332018-01-03
- FDAMagnetic Resonance System G-scan Brio (Part Number /Model No.101002000)2016-06-06
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.