FDAJuly 26, 2024device

Atellica IM CA 19-9 (250 Test)- In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN:10995489

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.

What to do

FDA enforcement status: Ongoing

Brands named

siemens healthcare diagnosticssiemenssiemens healthcare

UPCs

00630414598161

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.