FDAAugust 8, 2024device

Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed in 2008 Series Hemodialysis Machines with Part Numbers 190573, 190713, 190766, 190858, 190895, 190908, 191124, 191126, 191128, and 191130

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cuff.

What to do

FDA enforcement status: Ongoing

Brands named

fresenius medical carefreseniusfresenius medical