GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: a) 2110344-008; b) 2110344-009; c) 2110344-010; d) 2110344-011; e) 2110344-013; f) 2110344-019; g) 2110344-025; h) 2110344-029; i) 2110344-030; j) 2110344-032 ; Radiological image processing system

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Footprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the same day as the original report but at a different time, and only when launched on the following browsers. " Internet Explorer 10 (IE10) " Internet Explorer 11 (IE11) " Firefox.

What to do

FDA enforcement status: Ongoing