FDAJuly 26, 2018device

Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial portion of the plan.

What to do

FDA enforcement status: Terminated

Brands named

stryker