FDAAugust 29, 2025device

Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

What to do

FDA enforcement status: Ongoing

Brands named

getinge disinfection abgetingegetinge disinfection

UPCs