FDAApril 2, 2025device
Washer Disinfector Aquadis 56; Model Number: 56M
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
What to do
FDA enforcement status: Ongoing
Brands named
getinge disinfection abgetingegetinge disinfection
UPCs
07340153710351
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAGetinge 88-Series Washer- Disinfector. Model Number: 88-5.2025-08-29
- FDAWasher Disinfector Aquadis 56; Model Number: 56A2025-04-02
- FDAVaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.2025-01-15
- FDAMEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 720001B22024-10-03
- FDAMEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 710001B22024-10-03
- FDAMEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 720001F22024-10-03
- FDAMEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 710001B02024-10-03
- FDAMEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 700001F02024-10-03