FDASeptember 3, 2024device

GE Heathcare Giraffe OmniBed; neo natal incubator

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.

What to do

FDA enforcement status: Ongoing

Brands named

datex ohmedadatexdatex ohmeda