FDASeptember 3, 2024device
GE Heathcare Giraffe OmniBed Carestation, Model Number 2082844-001-XXX; neo natal incubator
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.
What to do
FDA enforcement status: Ongoing
Brands named
datex ohmedadatexdatex ohmeda
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAPanda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.2026-05-08
- FDAGiraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001, 2082663-002, M1241424 Product impacted only if service blender M1091607-R was installed during the last Preventative Maintenance2026-05-08
- FDAGiraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.2026-05-08
- FDAGE Heathcare Giraffe OmniBed; neo natal incubator2024-09-03
- FDAAespire 100. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.2024-07-12
- FDAAuxiliary Common Gas Outlet (ACGO) conversion/install kits sold under product names CONV KIT SCGO TO ACGO and ACGO ASSY FIELD INSTL. These kits are not affected specifically, however are included due to being purchased with or later used to convert to ACGO.2024-07-12
- FDAAisys CS2 with Et Control. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.2024-07-12
- FDAAespire 7100, 7100 (Variant), and Aespire 7100 S/5 Protiva configurations. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of genera...2024-07-12
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.