Tritanium TL Steerable Inserter, Cat. No. 48930000, UDI 07613327381900 Product Usage: The TL Steerable Inserter is a Class 1 instrument which is provided in a non-sterile condition. The inserter is designed to enter the disc space, with a fixed engagement between the inserter and the cage and the...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The firm received complaints related to the locking mechanism on the steerable inserter not functioning as intended. The firm confirmed that the issue is related to lack of lubrication of the instrument. There are two potential hazards associated with the implant pivoting unexpectedly during implant insertion into the disc space. First, if the implant is initially inserted into the disc space via the impaction method and the implant pivots unexpectedly, the implant potentially may unintentionally contact sensitive anatomy. Additionally, the user may potentially experience difficulty positioning the implant within the disc space due to the lack of rigidity. A third potential hazard is associated with the inserter not unlocking at the articulating pivot head during implant positioning within the disc space. Once the implant is in the prepared disc space and contacts the ventral annulus, the surgeon unlocks the articulating pivot head and positions the implant perpendicular t
What to do
FDA enforcement status: Terminated
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDATRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300.2026-05-19
- FDATRI TS BASEPLATE SIZE 4. Part Number: 5521-B-400.2026-05-19
- FDAMICS3 Angled Sagittal Saw Attachment; Part Number: 2104902026-02-02
- FDAThe Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in...2025-11-12
- FDA1. EXETER V40 STEM 44MM NO 2, Model/Catalog Number: 0580-1-442; 2. EXETER V40 STEM 37.5MM NO 0, Model/Catalog Number: 0580-1-352;2025-08-28
- FDATriathlon Universal TS Baseplate Size 2 - A single use sterile implantable device intended for cemented applications during primary and revision Total Knee Arthroplasty (TKA) when the Triathlon Knee System is selected for use. Catalog Number: 5521-B-2002024-09-04