FDAOctober 11, 2023device

Ventlab, LLC STAT-Check manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. STAT-Check Adult Resuscitator REF #: SC9101B; 2. STAT-Check Infant Resuscitator RES #s: SC7101B & SC7101B-M0; 3. STAT-Check Pediatric Resuscitator REF #: SC8121B.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

What to do

FDA enforcement status: Ongoing

Brands named

sunmed

UPCs

1088948310925030889483109254108894831606713088948316067510889483110041308894831100451088948309867730889483098671

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.