FDAOctober 11, 2023device

Curaplex manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Curaplex Adult Premium BVM with Bag Reservoir, Manometer, Filter, Peep Valve & Mask, REF # 2442-BVMPAD; 2. Curaplex Child Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter & Mas...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

What to do

FDA enforcement status: Ongoing

Brands named

sunmed

UPCs

0081007163312600812277037050008122770370740081227703708100812277037067

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Curaplex manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Curaplex Adult Premium BVM with Bag Reservoir, Manometer, Filter, Peep Valve & Mask, REF # 2442-BVMPAD; 2. Curaplex Child Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter & Mas... — Recall Details · AllClear