FDAOctober 1, 2025device
Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303, OCRD-SP, OCRD-MP, OCRD-UL; Product Description: In Vitro Diagnostic external control
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.
What to do
FDA enforcement status: Ongoing
Brands named
microgenics
UPCs
00884883011946008848830119530088488301196000884883011977008848830119840088488301199100884883012004
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAQMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-32022-01-20
- FDAThermo Scientific QMS Everolimus Assay; Model number: 0380000 (US Distribution); 0373852 (Foreign Distribution) Product Usage: The QMS Everolimus Assay is intended for the quantitative determination of Everolimus, in human whole blood on automated clinical chemistry analyzers. The results obtaine...2016-08-15
- FDADRI Salicylate Serum Tox Assay, Catalog number 0977. Toxicology: The DRI Salicylate Serum Tox Assay is intended for the quantitative determination of salicylate in human serum or plasma.2016-07-25
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →