FDAJanuary 20, 2022device

QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.

What to do

FDA enforcement status: Ongoing

Brands named

microgenics

UPCs

0884883012592

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3 — Recall Details · AllClear