FDAOctober 25, 2018device

QUADROX-i Neonatal with integrated arterial filter and VHK 11000 (VKMO 11000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055207 The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.

What to do

FDA enforcement status: Terminated

Brands named

maquet cardiovascular us salesmaquetmaquet cardiovascular