Blood Monitoring Unit (BMU 40), model no. 701040852 - Product Usage: is to monitor blood parameters during cardiopulmonary bypass (CPB) or similar procedures with extracorporeal circulation, which require continuous monitoring of the arterial and/or venous blood parameters: partial pressure of ox...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and proximal areas on the printed circuit board (PCB) where as the standard requires a creepage distance of 4 mm.
What to do
FDA enforcement status: Terminated
Brands named
maquet cardiovascular us salesmaquetmaquet cardiovascular
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- CPSCJC Sales Recalls Lil' Buddies Pet Laser Toys Due to Risk of Serious Injury or Death from Battery Ingestion Hazard; Violates Mandatory Standard for Consumer Products with Button Cell Batteries2026-04-16
- FDABubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;2026-01-09
- FDAThe product is a handheld ultraviolet-C germicidal wand and contains a narrow-diameter, tubular lamps with a peak emission wavelength in the 254 nm range.2025-08-29
- FDAHeartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.2025-08-15
- FDAHeartstring III Proximal Seal System. Intravascular anastomosis occluder.2025-08-15
- FDAHeartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.2025-08-15
- FDAVasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting T...2025-08-06
- FDACorin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.2025-07-14
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →