FDASeptember 24, 2018device

Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard

What to do

FDA enforcement status: Completed

Brands named

invivo

UPCs

45356466908100884838091474

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.