DynaCad software consist of an MR Analysis Server software and the viewer workstation software. The MR Analysis software consists of DynaCad Breast, DynaCad Prostate, and DynaCad Advanced PK for other MR analyses modules. Product Usage: Intended to be used as a post processing software package de...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Following update from 3.3 to 3.5 of the DynaCAD software it was noted that the Ktrans map was not rendered correctly on the DynaCad Client. The defect causes the pharmacokinetic (PK) color maps to display incorrectly when viewed from remote DynaCAD client computers and could result in visually underestimating calculated Ktrans, Kep, and iAUGC values. The defect also impacts DynaCAD s on-the-fly calculation of Apparent Diffusion Coefficient (ADC) maps. If the ADC values are computed on-the-fly by DynaCAD, the ADC values and colors will also display incorrectly if viewed on a remote DynaCAD client.

What to do

FDA enforcement status: Terminated

Brands named

invivo