Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W Product Usage: The Heater Cooler Unit (HCU) 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Failure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board. Board failure typically occurs when the compressor is switched on (prior to patient use); however, it can occur during patient use. This can shut down the compressor or the entire HCU 30 Type 2 unit.
What to do
FDA enforcement status: Terminated
Brands named
maquet cardiovascular us salesmaquetmaquet cardiovascular
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- CPSCJC Sales Recalls Lil' Buddies Pet Laser Toys Due to Risk of Serious Injury or Death from Battery Ingestion Hazard; Violates Mandatory Standard for Consumer Products with Button Cell Batteries2026-04-16
- FDABubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;2026-01-09
- FDAThe product is a handheld ultraviolet-C germicidal wand and contains a narrow-diameter, tubular lamps with a peak emission wavelength in the 254 nm range.2025-08-29
- FDAHeartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.2025-08-15
- FDAHeartstring III Proximal Seal System. Intravascular anastomosis occluder.2025-08-15
- FDAHeartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.2025-08-15
- FDAVasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting T...2025-08-06
- FDACorin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.2025-07-14
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →