FDADecember 8, 2015device
Perkin Elmer ClWIZARD2 5-detector, 550 samples Product Code: 2470-0050, 3470-0050. The Wizard2 gamma counter is intended to detect and count gamma radiation emitted by clinical samples.
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.
What to do
FDA enforcement status: Terminated
Brands named
perkinelmer
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAAutoDELFIA¿ hCG Kit, In Vitro Diagnostic for the determination of human gonadotropin (hCG) Product Code: B007-112 - Product Usage: Quantitative determination of human gonadotrophin (hCG).2019-01-10
- FDADELFIA¿ hCG Kit, In Vitro Diagnostic for the determination of human gonadotropin (hCG) Product Code: A007-101 - Product Usage: Quantitative determination of human gonadotrophin (hCG) in serum.2019-01-10
- FDAAutoDELFIA¿ hCG Kit, In Vitro Diagnostic for the determination of human gonadotropin (hCG) Product Code: B007-101 - Product Usage: Quantitative determination of human gonadotrophin (hCG).2019-01-10
- FDAPerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 with Hot Fix 2 Software intended to identify compounds in human samples by ionizing the compound and separating the resulting ions. Part Number: BC0048722018-09-24
- FDAPerkinElmer 226 Sample Collection Device2017-12-19
- FDASpecimen Gate Screening Center, Part Number 5002-0500, All software versions from 1.0 to 1.8 (current software version) Specimen Gate Screening Center is used for data management of neonatal screening test results and demographics by qualified laboratory personnel in newborn screening programs.2017-11-17
- FDANeoBase Succinylacetone Assay Solution, Product Number: 3042-0020; Affected UDI Numbers: (01)06438147295708(17)180531(10)654184 (01)06438147295708(17)180531(10)653414 (01)06438147295708(17)180531(10)653657 (01)06438147295708(17)180630(10)654923 (01)06438147295708(17)180630(10)655680 (01)064381472...2017-11-02
- FDAAutoDELFIA Neonatal immunoreactive trypsin(ogen) (IRT) kit, intended for the quantitative determination of human immunoreactive trypsin(ogen) (IRT) in blood specimens. (UDI: (01)06438147298716(17)190430(10)656479); Product Number: B005-2122017-09-25
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →