FDADecember 8, 2015device

Perkin Elmer WIZARD2 10-detector, 550 samples Product Code: 2470-0100, 3470-0100. The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.

What to do

FDA enforcement status: Terminated

Brands named