FDANovember 17, 2016device

Monaco TRP System The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Contours are not handled correctly if the CT dataset contains slices with DICOM Z coordinates of xx.x50mm.

What to do

FDA enforcement status: Terminated

Brands named

elekta

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.