FDAJanuary 29, 2015device

Elekta Precise Treatment Table

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may occur if a position error is not detected.

What to do

FDA enforcement status: Terminated

Brands named

elekta