FDAMarch 15, 2019device

Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Catalogue Number C2003-2, Lot #070318) IFU: "RAMP Cardiac Controls *** CAT. No.: C2003-2 *** Size: 3 x 3 ML". Product Usage: use in the quality control of cardiac marker assays run on the RAMP¿ p...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The form received three complaints related to the user obtaining a result that fell below the target range for CK-MB provided in the RAMP Cardiac Controls Instructions for Use. Via in-house monitoring, determined that liquid Cardiac Control has seen a decrease in CK-MB Levels of up to 57% signal loss within 11 months. Shelf-life of product is 24 months.

What to do

FDA enforcement status: Terminated

Brands named

response biomedicalresponse

UPCs

10627966005084