CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment

What to do

FDA enforcement status: Ongoing