CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAAccess Total T4 Calibrator, Catalog No. 338052026-03-16
- FDADxC 700 AU, REF: B86444, B864462026-01-30
- FDACellMek SPS Sample Preparation System, REF: C44603, with software2025-12-23
- FDAThe LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end prod...2025-12-08
- FDABeckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit2025-11-26
- FDABeckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit2025-11-26
- FDACOULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemake...2025-11-10
- FDAAccess 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibration samples on the Access 2 Immunoassay system, Catalog Number 819012025-11-07
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →