FDAApril 27, 2023device

Intelli-C, Right side suspended Tabletop, REF: 03400000

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.

What to do

FDA enforcement status: Ongoing

Brands named

nrt x ray a snrt

UPCs

05713464000527