FDAApril 27, 2023device

Intelli-C, Left side suspended Tabletop, REF: 03400010

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.

What to do

FDA enforcement status: Ongoing

Brands named

nrt x ray a snrt

UPCs

05713464000121

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Intelli-C, Left side suspended Tabletop, REF: 03400010 — Recall Details · AllClear