FDAMay 18, 2016device

Monaco RTP System Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

When DICOM exporting a 3D Monaco plan and the "Composite Field Sequencing" (CFS) checkbox is selected, the Y jaws will snap to the port shape when they should remain where they were defined.

What to do

FDA enforcement status: Terminated

Brands named

elekta

Recall history