FDAApril 2, 2026device

Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.

What to do

FDA enforcement status: Ongoing

Brands named

cook

UPCs

00827002473089

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.