FDAApril 2, 2026device

Beacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20 G31214 NR5.0-35-100-P-10S-PIG-CSC-20 G31216 NR5.0-35-70-P-10S-VCF-CSC-20 G31220 NR5.0-35-100-P-10S-VCF-CSC-20 G31221 Sizing catheters are single-use, sterile catheters with gold marker band...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.

What to do

FDA enforcement status: Ongoing

Brands named

cook

UPCs

0082700231213500827002312142008270023121660082700231220300827002312210

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.