FDAApril 7, 2020device

Aptio Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00) - Product Usage: is a modular system designed to automate Pre-Analytical and Post-Analytical process...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.

What to do

FDA enforcement status: Terminated