FDAMay 31, 2024device

The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches....

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.

What to do

FDA enforcement status: Ongoing

Brands named

maquet cardiovascularmaquet

UPCs

00607567700345

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches.... — Recall Details · AllClear