FDAJanuary 15, 2019device
Neo-Vent Infant T-Piece Resuscitation Circuit;Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0415, Qty:10, Length: 60" (1.5cm), Rx Only.
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.
What to do
FDA enforcement status: Terminated
Brands named
westmed
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDA1) Flex Extenders, 5.2" x 15mm OD x 15mm ID REF S6411P 2) Flex Extender, 5.2" x 15mm OD x 22/15mm ID REF 7 5132 The Westmed Flex Extenders are non invasive, single use/disposable, large bore, corrugated plastic tubes that provide a flexible, low resistance lightweight connection between breathing...2024-09-27
- FDAWestmed DISPOSABLE RESPIRATORY PRODUCTS NON-REBREATHER MASK, REF 3228-e, RX Only, CE0482, (01)10709078008020 - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.2020-11-11
- FDAREF 0562K (Adult Elongated O2/CO2 Mask - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.2020-11-11
- FDAREF 0562, Adult Elongated O2/CO2 Mask, 10'X0.50 GSL w/ Swivel Elbow, 2" Pigtail w/Female Luer Connectors, 7' Oxygen Supply Tubing, Rx Only, CE0482, (01)00709078001697 - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.2020-11-11
- FDAREF 0930 Adult Elongated CO2/O2 Mask, 7' Tubing w/ Threaded Nut, Sivel GSE, 10' GSL X .038" ID X .107" OD, Reflective Luer Connectors, Rx Only, CE0482, (01)00709078005718 - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.2020-11-11
- FDAREF 3102-E, Elongated Oxygen Mask, Adult Oxygen Elongated Mask with 8' Tubing, Rx Only, CE0482, (01)00709078003172 - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.2020-11-11
- FDAREF 0562F, Adult CO2/O2 Mask, 10'X0.50 GSL w/ Swivel Elbow, 2" Pigtail w/Female Luer Connectors, 7' Oxygen Supply Tubing, 0.8 Micron Filter, Rx Only, CE0482, (01)00709078005664 - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.2020-11-11
- FDANeo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask; Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0420, Qty:10, Length: 60" (1.5cm), Rx Only.2019-01-15