FDAJune 19, 2017device
Nova Biomedical Stat Profile Prime Analyzer as follows: PN 52855 Stat Profile Prime ABG Analyzer PN 53655 Stat Profile Prime ABG Analyzer with Scanner PN 52856 Stat Profile Prime CCS Analyzer PN 53656 Stat Profile Prime CCS Analyzer with Scanner PN 53657 Stat Profile Prime CCS Comp Analyzer with ...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Software- scheduled A-QC analysis could initiate prior to a patient result being released by the operator and result in the pending patient result being overwritten with the A-QC test results.
What to do
FDA enforcement status: Terminated
Brands named
nova biomedicalnova
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDASparrow Ascent Patient Controller, REF's: 11-110-K, 11-310-K, 11-320-K, 11-321-K, 11-310-K-28, 11-320-K-28, 11-321-K-282025-07-02
- FDABigfoot Unity Diabetes Management System SW-300689-01 Bigfoot Inject v1 Pen Cap LCAP Firmware 1.3.0 Release Configuration, Rev H SW-300689-02 Bigfoot Inject v1 Pen Cap RCAP Firmware 1.3.0 Release Configuration, Rev H The Bigfoot Unity Diabetes Management System that include the Pen Caps (Rapid-Ac...2025-05-21
- FDACenterline Biomedical IOPS Guidewire 2, REF ATW-2; Sensorized wire used to navigate through vasculature to facilitate placement of a catheter2025-04-11
- FDAStatStrip Glucose Ketone Hospital Meter System. Model/Catalog Number: 63683. Prescription use blood glucose ketone meter for near-patient testing2024-11-20
- FDAStatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-patient testing.2024-11-20
- FDAStatStrip Glucose Ketone (mmol/L) Hospital Meter System. Model/Catalog Number: 63910. Prescription use blood glucose ketone meter for near-patient testing2024-11-20
- FDATriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509)2024-11-11
- FDABrand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR PURPLE Model/Catalog Number: 4329PL Product Description: ROLLING WALKER MOBILITY AID2024-08-26
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →