FDASeptember 11, 2025device

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

What to do

FDA enforcement status: Ongoing

Brands named

insulet

UPCs

10385083000459

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.