FDAOctober 17, 2022device

18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: Assembly, Dash Final PDM U100, mmol/L PT-000030: Assembly, DASH PDM, Canada Remote Controller (Personal Diabetes Manager or PDM) associated with Insulin Pump (Pod) *The part numbers ab...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.

What to do

FDA enforcement status: Terminated

Brands named

insulet

UPCs

10385082000009103850820001221038508200013910385082000146

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.