FDAFebruary 1, 2021device

Omnipod DASH Insulin Management System User Guide (OUS Only) Product Catalog Number: PT 000002 XXX XXX XX PT 000318 PT 000319 Distributed to: Canada (French only), the Netherlands, France, Austria, Germany, Belgium, Greece, Turkey, Norway, Sweden, Denmark, and Finland

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Certain foreign user guides include a misprint which could lead to an incorrect bolus delivery and hypoglycemia. Only Outside the US (OUS): Error in the Omnipod DASH User Guide text, page 6 of the User Guide, the text states that the comma (,) is not used for the decimal on the PDM, when in fact it is the period (.) that is not used.If a user were to use the period as a decimal separator when entering a bolus value, this decimal separator would not be recognized by the device and any numbers entered after the period could result in a larger value than intended

What to do

FDA enforcement status: Terminated

Brands named

insulet