FDAJune 6, 2023device

GE Healthcare Giraffe Omnibed Carestation CS1, combination infant incubator and infant warmer. Model Numbers: 1) 2082844-001-01040418; 2) 2082844-001-01046784; 3) 2082844-001-01052324; 4) 2082844-001-01052572; 5) 2082844-001-01055337; 6) 2082844-001-01065693; 7) 2082844-001-01069008; 8) 2082844-0...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.

What to do

FDA enforcement status: Completed

Brands named

datex ohmedadatexdatex ohmeda

UPCs

00840682116862008406821168