FDASeptember 28, 2021device

Flexor Check-Flo Introducer, Introducer Set, RPN KCFW-6.0-35-45-RB, GPN G09908 The introducer sheath incorporates a hydrophilic coated Flexor shaft with a hemostasis valve and is provided with a dilator.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Affected lots may be manufactured incorrectly with the radiopaque marker band located just below the Check-Flo proximal fitting instead of the distal tip. The issue may not be identified until the device is placed in the patient and viewed under fluoroscopy, leading to increased procedural time.

What to do

FDA enforcement status: Terminated

Brands named

cook

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Flexor Check-Flo Introducer, Introducer Set, RPN KCFW-6.0-35-45-RB, GPN G09908 The introducer sheath incorporates a hydrophilic coated Flexor shaft with a hemostasis valve and is provided with a dilator. — Recall Details · AllClear