FDASeptember 19, 2025device

DxI 9000 Access Immunoassay Analyzer, Part Number C11137

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.

What to do

FDA enforcement status: Ongoing

Brands named

beckman coulterbeckman

UPCs

15099590732103

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →