FDAMay 24, 2018device

Software Package, Sorting-Drive 4.2 User Kit, Part No. B73488, Manufactured by: Beckman Coulter Biomedical GmbH The Sorting-Drive itself is an optional software to be used with the Automate 2500 Family of products. The system automates the process of sorting, decapping, and archiving samples. The...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Sorting-Drive software version 4.2 (released in October 2015) introduced a new option which included the Sorting-Drive Client that is used for manual printing of barcodes. When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports this does not affect automated label creation or reprinting of individual labels and that no other system functions are affected.

What to do

FDA enforcement status: Terminated

Brands named

beckman coulterbeckman