FDADecember 1, 2015device

UniCel DxH 800 Coulter Cellular Analysis System, Software Version 3.0.2.0, Part No. 629029, B24465, B24802 The UniCel DxH 800/DxH 600 analyzers are quantitative, multi-parameter, automated hematology analyzers for in vitro diagnostic use in screening patient populations found in clinical laborato...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen identification (ID) but different Patient identification when manually editing pending orders at the System Manager.

What to do

FDA enforcement status: Terminated

Brands named

beckman coulterbeckman

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.