FDADecember 1, 2015device
UniCel DxH 600 Coulter Cellular Analysis System, Software Version 3.0.2.0, Part No. 775222 The UniCel DxH 800/DxH 600 analyzers are quantitative, multi-parameter, automated hematology analyzers for in vitro diagnostic use in screening patient populations found in clinical laboratories. The DxH 80...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen identification (ID) but different Patient identification when manually editing pending orders at the System Manager.
What to do
FDA enforcement status: Terminated
Brands named
beckman coulterbeckman
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAAccess Total T4 Calibrator, Catalog No. 338052026-03-16
- FDACHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B232812026-01-30
- FDADxC 700 AU, REF: B86444, B864462026-01-30
- FDACellMek SPS Sample Preparation System, REF: C44603, with software2025-12-23
- FDAThe LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end prod...2025-12-08
- FDABeckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit2025-11-26
- FDABeckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit2025-11-26
- FDACOULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemake...2025-11-10
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.