UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended use is as follows: The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean micr...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

When the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review. Erroneous results may be reported from the laboratory which may trigger unnecessary or inappropriate patient treatment.

What to do

FDA enforcement status: Terminated

Brands named

beckman coulterbeckman

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended use is as follows: The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean micr... — Recall Details · AllClear